Common Subsystems
For more information on Common Subsystems, please refer to the configured COMT# Template SFS.
After Vacuum Flag
-
Turns On:
-
(Following EO Inject or EO dwell phase whenever Vacuum Pump is ON) AND (estimated EO concentration OR when available measured EO concentration is greater than High Concentration End Limit).
-
-
Turns Off:
-
When the estimated EO concentration AND when available measured EO concentration are less than High Concentration End Limit.
-
Cycle Complete
Turns ON at the end of the cycle when the door is opened and chamber enters dormant.
Cycle in Progress
-
Turns On:
-
When a cycle is started.
-
-
Turns Off:
-
(Door interlock is satisfied AND End Phase AND Not Dormant) OR Dormant.
-
Emission Control Request
The emission control request output will tell the emission control system to prepare to receive EO flow from the vacuum pump.
-
Turns On:
-
So many seconds (emission control lead time parameter 55) before the end of EO Dwell OR
-
During the first Vacuum phase occurring after EO Inject when the EO concentration is greater than 13.5 mg/l OR
-
If high pressure alarm occurs during EO Dwell.
-
-
Turns Off:
-
High Concentration End Limit (Parameter 56) is NOT set and in the last Air In-bleed phase OR
-
When estimated EO concentration is less than High Concentration End Limit (Parameter 56) OR
-
Chamber is in Dormant sequence.
-
User OK
-
OFF State: If enabled (i.e. equipment configuration “User OK Hold Enabled” Keyword 15 is set to true), an event will be raised and the beacon will be activated; the chamber will go into a mode according to event configuration.
-
ON state: When User OK signal comes back on, or is no longer enabled (i.e. Keyword 15 is set to false), an event will be raised and the beacon will be activated; the chamber will remain in the previous mod.e
Remote Hold
If Remote Hold signal turns OFF while Emission Control Output is ON, the following will occur:
-
Chamber goes on Hold if “Remote Hold Enabled” (Keyword ID 2) is true.
-
Alarm 318 will be raised: “Remote Hold signal turned off while Remote Hold Enabled option was true”.
-
-
If “Remote Hold Enabled” (Keyword ID 2) is false:
-
Alarm 96 will be raised: “Remote Hold signal has turned off”.
-
-
If in Vacuum phase then:
-
Vacuum block valve closes.
-
Any vacuum pump motor stays on.
-
Anti-cavitation block valve stays on.
-
If Remote Hold signal turns OFF while Emission Control Output is OFF, there is no impact on chamber operation.
Remote Hold signal turns ON:
-
If the Emission Control Output is OFF, it will have no impact on chamber operation.
-
If the Emission Control Output is ON and no other Hold occurred, chamber will resume normal operation.
-
Alarm 319 will be raised: “Remote Hold signal turned on while Emission Control output is on.”
-
Building Power
The system will monitor the presence of building power.
Operation
-
If the system detects absence of building power, all analog and discrete outputs will turn off after a 60 second delay.
-
Chamber will go into Stop mode.
-
When the condition is reset, normal product process can resume after a 10 second delay.
Emergency Stop
The system will monitor the emergency stop condition.
Operation
-
If emergency stop is activated, all analog and discrete inputs will turn off.
-
Chamber will go into Stop mode.
-
A hardwired reset button will reset the emergency stop condition after the Emergency stop button has been pulled back to the normal position.
-
When the condition is reset, normal product process can resume after a 10 second delay.